Safety and Risk-Benefit Profile of Microbubble Contrast Agents in Echocardiography

Register or Login to View PDF Permissions
Permissions× For commercial reprint enquiries please contact Springer Healthcare: ReprintsWarehouse@springernature.com.

For permissions and non-commercial reprint enquiries, please visit Copyright.com to start a request.

For author reprints, please email rob.barclay@radcliffe-group.com.
Average (ratings)
No ratings
Your rating

Abstract

In October 2007, the US Food and Drug Administration (FDA) imposed a ‘black box’ warning, new disease state contraindications, and a mandatory 30-minute monitoring period on the ultrasound contrast agents Definity® (perflutren lipid microspheres, Lantheus Medical Imaging, North Billerica, MA) and Optison™ (perflutren protein-type A microspheres, GE Healthcare, Buckinghamshire, UK), effectively restricting their use. These labeling changes were prompted by spontaneous reports of four deaths and ~190 other serious cardiopulmonary reactions that occurred in temporal relation to contrast agent injection. Many healthcare providers were critical of the FDA action, citing the previously documented excellent safety profile of these agents, the possible confounding effect of ‘pseudocomplication,’ and the known increased risk associated with the more invasive testing that would be necessary in many patients. Since then, several large-scale ultrasound contrast agent safety studies have been completed, including ~200,000 patients. In this article, we review the efficacy of ultrasound contrast agents, implications of the initial FDA action, newly published safety data, and more recent changes to the product labeling for both agents following an FDA Cardio-Renal Advisory Committee meeting in June 2008.

Disclosure:Michael L Main,MD, has received research grants from Lantheus Medical Imaging and GE Healthcare, and has served as a consultant for Acusphere, Inc. Taiyeb M Khumri, MD, has no conflicts of interest to declare.

Received:

Accepted:

Support:Supported by Lantheus Medical Imaging. The views expressed are those of the authors and not necessarily those of Lantheus Medical Imaging.

Correspondence Details:Michael L Main, MD, Cardiovascular Consultants, 4330 Wornall Road, Suite 2000, Kansas City, MI 64111. E: mmain@cc-pc.com

Copyright Statement:

The copyright in this work belongs to Radcliffe Medical Media. Only articles clearly marked with the CC BY-NC logo are published with the Creative Commons by Attribution Licence. The CC BY-NC option was not available for Radcliffe journals before 1 January 2019. Articles marked ‘Open Access’ but not marked ‘CC BY-NC’ are made freely accessible at the time of publication but are subject to standard copyright law regarding reproduction and distribution. Permission is required for reuse of this content.

Ultrasound contrast agents capable of transpulmonary passage following intravenous injection have been commercially available since the 1990s. These agents are microbubbles, which are smaller than red blood cells and persist long enough (due to reduced rate of gas diffusion) to reach the left ventricle (LV).1–3 Ultrasound contrast agent applications included LV cavity opacification, enhancement of spectral Doppler signals and evaluation of myocardial perfusion at rest or post-stress.3 The commercially available second-generation ultrasound agents are approved by the US Food and Drug Administration (FDA) only for the enhancement of LV endocardial delineation in patients with baseline suboptimal examinations.3 There are currently two commercially available ultrasound contrast agents in the US: Optison™ (perflutren protein-type A microspheres, GE Healthcare, Buckinghamshire, UK) and Definity® (perflutren lipid microspheres, Lantheus Medical Imaging, North Bellirica).

Efficacy of Ultrasound Contrast Agents

Although tissue harmonic imaging has significantly enhanced the diagnostic quality of baseline transthoracic echocardiography,4 the increasing prevalence of obesity and lung disease has added an extra layer of challenge; at least 10–15% of echocardiograms are technically difficult or non-diagnostic despite the use of harmonic imaging. The proportion of non-diagnostic studies is significantly increased during stress echocardiography and in echocardiograms performed in the intensive care unit.3

In non-selected patients, contrast-enhanced echocardiography leads to more accurate assessment of LV volumes and ejection fraction compared with magnetic resonance imaging.5,6 Even with the use of harmonic imaging, ultrasound contrast enhances accuracy and reproducibility in calculation of LV systolic function.6 In selected patients with excellent echocardiographic windows and adequate endocardial visualisation of all segments with harmonic imaging, the utilisation of contrast echocardiography reduces inter- and intra-reader variability in the assessment of LVEF.7

In patients with abnormal LV systolic function undergoing evaluation for LV remodeling, the use of ultrasound contrast agents reduces reader variability and increases accuracy compared with harmonic imaging with computed tomography as a reference standard.8 In intensive care patients, the use of ultrasound contrast agents increases the diagnostic yield of both individual myocardial segment evaluation and overall LV function assessment, and should be considered in all patients with non-diagnostic studies.9

Contrast echocardiography also improves segmental evaluation during dobutamine stress echocardiography10 and in mechanically ventilated patients,11 leading to an increased diagnostic yield. Additionally, contrast-enhanced echocardiography reduces the downstream utilisation of alternative testing resources, making ultrasound contrast agents cost-effective for rest and stress imaging.12–15

Adverse Effects and Contraindications Based on Pre-marketing Studies

In pre-marketing studies, Optison™ was administered to 279 patients.16 Forty-seven patients (16.8%) reported at least one adverse event. Of these, one event was serious and required treatment with antihistamines for hypersensitivity manifestations of dizziness, nausea, flushing and temperature elevation. No deaths were reported during the clinical studies. Headache (5.4%) was the most commonly reported side effect. One thousand, seven hundred and sixteen patients received Definity in pre-marketing clinical trials.17 Of these, 144 (8.4%) had at least one treatment-related adverse reaction. There were eight deaths, which were attributed to underlying disorders. There were 11 other serious adverse events, including one hypersensitivity reaction with urticaria and pruritus; other patients experienced dizziness, chest pain, dyspnoea or back pain. Headache (2.3%) and back and renal pain (1.2%) were the most common side effects, and resolved spontaneously without treatment.

The use of ultrasound contrast agents is contraindicated in patients with right-to-left, bi-directional or transient right-to-left cardiac shunts that can allow microspheres to bypass the pulmonary particle-filtering mechanisms and directly enter the arterial circulation, potentially resulting in microvascular occlusion and ischaemia. These agents should also not be injected intra-arterially or in patients who have a known hypersensitivity to perflutren. The use of Optison is also contraindicated in patients with an allergy to egg (albumin).

US Food and Drug Administration Boxed Warning – October 2007

In October 2007, the US FDA mandated labelling revisions for both commercially available perflutren-containing ultrasound contrast agents largely based on four reported deaths that were temporally related (but not clearly causally attributable to) ultrasound contrast agent injection. The labelling revisions included a ‘Boxed Warning’ highlighting the potential for “serious cardiopulmonary reactions”, new disease state contraindications including worsening or clinically unstable heart failure, acute myocardial infarction or acute coronary syndrome, serious ventricular arrhythmia or high risk for arrhythmia based on QT interval prolongation, and respiratory failure, severe emphysema, pulmonary emboli or other conditions that cause pulmonary hypertension. The FDA also mandated a 30-minute monitoring period for all patients following contrast agent injection.

Implications of the US Food and Drug Administration Warning

The FDA warnings and contraindications effectively restricted the use of contrast echocardiography in acutely ill patients; both published data and clinical experience indicated that these were the patients who were most likely to derive diagnostic benefit from the addition of an ultrasound contrast agent. Immediately following the October 2007 FDA product labelling changes, use of ultrasound contrast agents in the US decreased dramatically, with many medical centres discontinuing contrast echocardiography altogether.

Many echocardiographers were immediately critical of the FDA action, noting that there was no direct evidence for causality in the deaths reported in close temporal association to ultrasound contrast agent injection. Experts also noted the previously documented excellent safety profile of the commercially available agents, the possible confounding effect of ‘pseudocomplication’ (death or other adverse outcome due to progression of underlying disease and not diagnostic testing or therapeutic intervention) and the known increased risk of other more invasive testing (such as transoesophageal echocardiography), which would now be necessary in many patients.18,19 Within the past year, several large-scale safety studies have been published that have objectively defined the safety profile of ultrasound contrast agents in nearly 200,000 patients.

Recently Published Safety Data

In a research letter published in the Journal of the American Medical Association, Herzog reported on 16,025 patients on the Hennepin County Medical Center database who received an ultrasound contrast agent. Optison was used in 3,051 studies and Definity was used in 12,974 patients. A total of 20 adverse events episodes were recorded, with four serious adverse events (serious adverse event rate 0.03%) and no fatalities. Many of the adverse events (including the four serious events) were consistent with allergic-type phenomena; in the patients experiencing these events, 65% were women and 65% had a previous history of atopy.20

Kusnetzky et al. reported a large retrospective analysis of mortality in hospitalised patients undergoing echocardiography with (6,196 patients) and without (12,475 patients) an ultrasound contrast agent at Saint Luke’s Hospital System in Kansas City over a three-year period. Twenty-four-hour mortality rates were 0.42% for patients receiving Definity and 0.37% for patients undergoing unenhanced echocardiography (p=NS). Despite no increased mortality risk in the contrast echocardiography group, these patients exhibited more significant co-morbidities and illness severity than those undergoing non-contrast echocardiography.21

In a large multicentre, retrospective analysis conducted by the American Society of Echocardiography (ASE),22 data on 78,383 contrast-enhanced studies were reported (66,164 Definity and 12,219 Optison), including ~10,000 critically ill patients hospitalised in intensive care units. In this study, there were only eight serious adverse events (0.01%), four of which were serious and secondary to an anaphylactoid reaction (0.006%); none of these occurred in the critically ill. There were no deaths attributable to ultrasound contrast agent use. These data were consistent with the very low serious adverse event rate in the Hennepin County database.20

Main et al. utilised the Premier Perspective Database, the nation's largest inpatient drug utilisation database,23 to analyse mortality in 4,300,966 patients who underwent inpatient echocardiography with 58,254 contrast-enhanced studies.24 Short-term mortality rates (death on the day of or day following echocardiography) were similar in patients undergoing echocardiography with or without the use of an ultrasound contrast agent (1.06 versus 1.08%; p=0.613). Multivariable logistic regression analysis (correcting for case mix and clinical co-variates) revealed that patients receiving Definity were 24% less likely to die acutely compared with patients undergoing unenhanced echocardiography.24

A third, retrospective, multicentre study performed at Saint Louis University, the University of Nebraska and the Mayo Clinic (data acquired between 1999 and 2007) evaluated a total of 42,408 consecutive patients who received an ultrasound contrast agent. In 18,749 patients who underwent stress echocardiography with contrast, there were no short-term (30-minute) adverse events attributable to contrast, and 30-day myocardial infarction and mortality rates were similar in patients receiving a contrast agent compared with a control group. In 23,659 patients who underwent resting contrast studies with short-term follow-up, there were no contrast-related serious adverse events at 30 minutes and no difference in myocardial infarction or mortality rates at 24 hours compared with a control group who did not receive a contrast agent.25

Shaikh reported on 5,069 patients who underwent the stress echocardiography procedure; an ultrasound contrast agent was administered to 2,914 of these patients. A higher proportion of both inpatients and patients undergoing dobutamine stress had contrast-enhanced studies. Despite the fact that patients receiving an ultrasound contrast agent tended to be older and were more likely to have reduced LV systolic function, there were no deaths in the group receiving contrast. One uncomplicated acute myocardial infarction and one anaphylactoid reaction occurred in the contrast agent group (p=0.51). Thus, while ultrasound contrast agents were more likely to be used in older hospitalised patients with depressed LV systolic function, there was no increase in the rates of major adverse events.26

A second study evaluating the safety of ultrasound contrast agents during stress echocardiography was reported by the Cleveland Clinic. A series of 4,786 patients who underwent contrast-enhanced stress echocardiograms (2,022 with dobutamine stress and 2,764 with exercise stress) were compared with a control group of 5,012 patients matched for test year and type who did not receive an ultrasound contrast agent. There was no difference in serious adverse event rate, death within 24 hours, cardiac arrest or sustained ventricular tachycardia between the contrast and non-contrast groups.27

Based on these new data (either published or in press in 2008), it is reasonable to conclude that: serious adverse events consistent with anaphylactoid reactions occur in ~1:10,000 patient doses; there is no increased mortality in hospitalised patients undergoing contrast echocardiography compared with patients undergoing echocardiography without an ultrasound contrast agent, despite the fact that patients receiving an ultrasound contrast agent exhibit more co-morbidities and higher acuity of illness; and contrast administration is safe in stress echocardiography.

Subsequently, through FDA review of additional information received after October 2007, and in accordance with the findings of an FDA Cardio-Renal and Safety Advisory Committee meeting on 24 June 2008, the FDA determined that, in some patients, the benefits from the diagnostic information that could be obtained through the use of Definity or Optison may outweigh the risks for serious cardiopulmonary reactions, even among those patients at particularly high risk for these reactions.28,29 Although a ‘Boxed Warning’ continues to highlight the potential for “serious cardiopulmonary complications”, all of the disease state contraindications issued in October 2007 have been revised to ‘warnings’, and the post-contrast 30-minute monitoring period now applies only to patients with pulmonary hypertension and unstable cardiopulmonary conditions. The FDA also recently announced that both manufacturers (Lantheus Medical Imaging and GE Healthcare) have agreed to conduct two studies designed to further evaluate the safety of Definity and Optison that will include a large retrospective study using a hospital database to determine acute mortality in hospitalised patients undergoing echocardiography with and without an ultrasound contrast agent, and an invasive haemodynamic study to evaluate for changes in pulmonary artery pressure and resistance immediately following contrast agent administration.30

Conclusions

The ‘Boxed Warning’ disease state contraindications and mandated 30-minute monitoring period issued by the FDA in October 2007 had a chilling effect on ultrasound contrast agent use in the US. Seven recently published studies have shown that: serious adverse events consistent with anaphylactoid reactions occur in ~1:10,000 patient doses; there is no increased mortality in hospitalised patients undergoing contrast echocardiography compared with patients undergoing echocardiography without an ultrasound contrast agent, despite the fact that patients receiving an ultrasound contrast agent exhibit more co-morbidities and higher acuity of illness; and contrast administration is safe in stress echocardiography. Based on these data and other considerations, the FDA recently revised product labelling for both agents, and clinicians can once again perform contrast echocardiography in those patients most likely to derive diagnostic benefit. Although these agents are ‘safe’, physicians and sonographers performing contrast echocardiography should be aware of the potential for a rare anaphylactoid reaction, and should be familiar with and have access to medications (most importantly epinephrine) to treat allergic-type reactions.

References

  1. Feinstein SB, Cheirif J, Ten Cate FJ, et al., Safety and efficacy of a new transpulmonary ultrasound contrast agent: initial multicenter clinical results, J Am Coll Cardiol, 1990;16:316–24.
    Crossref | PubMed
  2. Geny B, Mettauer B, Muan B, et al. Safety and efficacy of a new transpulmonary echo contrast agent in echocardiographic studies in patients, J Am Coll Cardiol, 1993;22:1193–8.
    Crossref | PubMed
  3. Main ML, Grayburn PA, Clinical applications of transpulmonary contrast echocardiography, Am Heart J, 1999;137:144–53.
    Crossref | PubMed
  4. Malhotra V, Nwogu J, Bondmass MD, et al., Is the technically limited echocardiographic study an endangered species? endocardial border definition with native tissue harmonic imaging and Optison contrast: a review of 200 cases, J Am Soc Echocardiogr, 2000;13:771–3.
    Crossref | PubMed
  5. Hundley WG, Kizilbash AM, Afridi I, et al., Administration of an intravenous perfluorocarbon contrast agent improves echocardiographic determination of left ventricular volumes and ejection fraction: comparison with cine magnetic resonance imaging, J Am Coll Cardiol, 1998;32:1426–32.
    Crossref | PubMed
  6. Malm S, Frigstad S, Sagberg E, et al., Accurate and reproducible measurement of left ventricular volume and ejection fraction by contrast echocardiography: a comparison with magnetic resonance imaging, J Am Coll Cardiol, 2004;44:1030–35.
    Crossref | PubMed
  7. Nayyar S, Magalski A, Khumri TM, et al., Contrast administration reduces interobserver variability in determination of left ventricular ejection fraction in patients with left ventricular dysfunction and good baseline endocardial border delineation, Am J Cardiol, 2006;98:1110–14.
    Crossref | PubMed
  8. Thomson HL, Basmadjian AJ, Rainbird AJ, et al., Contrast echocardiography improves the accuracy and reproducibility of left ventricular remodeling measurements: a prospective, randomly assigned, blinded study, J Am Coll Cardiol, 2001;38:867–75.
    Crossref | PubMed
  9. Reilly JP, Tunick PA, Timmermans RJ, et al., Contrast echocardiography clarifies uninterpretable wall motion in intensive care unit patients, J Am Coll Cardiol, 2000;35:485–90.
    Crossref | PubMed
  10. Porter TR, Xie F, Kricsfeld A, et al., Improved endocardial border resolution during dobutamine stress echocardiography with intravenous sonicated dextrose albumin, J Am Coll Cardiol, 1994;23:1440–43.
    Crossref | PubMed
  11. Kornbluth M, Liang DH, Brown P, et al., Contrast echocardiography is superior to tissue harmonics for assessment of left ventricular function in mechanically ventilated patients, Am Heart J, 2000;140:291–6.
    Crossref | PubMed
  12. Shaw LJ, Dittrich HC, Use of intravenous Optison contrast echocardiography reduces downstream resource use and enhances cost savings, Acad Radiol, 1998;(Suppl. 1):S250–51, discussion S252–3.
    Crossref | PubMed
  13. Shaw LJ, Gillam L, Feinstein S, et al., Use of an intravenous contrast agent (Optison) to enhance echocardiography: efficacy and cost implications. Optison Multicenter Study Group, Am J Manag Care, 1998;4:SP169–76.
    PubMed
  14. Shaw LJ, Impact of contrast echocardiography on diagnostic algorithms: pharmacoeconomic implications, Clin Cardiol, 1997;20:I39–48.
    Crossref | PubMed
  15. Thanigaraj S, Nease RF Jr, Schechtman KB, et al., Use of contrast for image enhancement during stress echocardiography is cost-effective and reduces additional diagnostic testing, Am J Cardiol, 2001;87:1430–32.
    Crossref | PubMed
  16. Labeling for Optison from Drugs@FDA, 2008. Available at: www.fda.gov/cder/foi/label/2008/020899s011lbl.pdf
  17. Labeling for Definity from Drugs@FDA, 2008. Available at: www.fda.gov/cder/foi/label/2008/021064s009lbl.pdf
  18. Keen C, Ultrasound contrast advocates take aim at FDA black box warning, 2008. Available at: www.auntminnie.com/index.asp? Sec=sup&Sub=car&Pag=dis&ItemId=79499
  19. Main ML, Goldman JH, Grayburn PA, Thinking outside the ‘box’—the ultrasound contrast controversy, J Am Coll Cardiol, 2007;50:2434–7.
    Crossref | PubMed
  20. Herzog CA, Incidence of adverse events associated with use of perflutren contrast agents for echocardiography, JAMA, 2008;299:2023–5.
    Crossref | PubMed
  21. Kusnetzky LL, Khalid A, Khumri TM, et al., Acute mortality in hospitalized patients undergoing echocardiography with and without an ultrasound contrast agent: results in 18,671 consecutive studies, J Am Coll Cardiol, 2008;51:1704–6.
    Crossref | PubMed
  22. Wei K, Mulvagh SL, Carson L, et al., The Safety of Definity and Optison for Ultrasound Image Enhancement: A Retrospective Analysis of 78,383 Administered Contrast Doses, J Am Soc Echocardiogr, 2008 (Epub ahead of print).
  23. PREMIER Research Services, 2008. Available at: www.premierinc.com/quality-safety/tools-services/prs/index.jsp
  24. Main ML, Ryan AC, Davis TE, et al., Acute mortality in hospitalized patients undergoing echocardiography with and without an ultrasound contrast agent; multicenter results in 4,300,966 consecutive patients, Am J Cardiol, 2008; in press.
  25. Dolan MS, Gala S, Dodla S, et al., Safety and efficacy of commercially available ultrasound contrast agents for rest and stress echocardiography: a multicenter experience, J Am Coll Cardiol, 2008; in press.
  26. Shaikh K, Chang SM, Peterson L, et al., Safety of Contrast Administration for Endocardial Enhancement during Stress Echocardiography Compared With Non-contrast Stress, Am J Cardiol, 2008; in press.
  27. Gabriel RS, Smyth YM, Menon V, et al., Safety of Ultrasound Contrast Agents in Stress Echocardiography, Am J Cardiol, 2008; in press.
  28. Summary minutes of the Cardiovascular and Renal Drugs Advisory Committee 2008. Available at: www.fda.gov/ohrms/dockets/ac/08/minutes/2008-4369m1-final
  29. United States Food and Drug Administration Center for Drug Evaluation and Research alert for healthcare providers, 2008. Available at: www.fda.gov/cder/drug/infopage/microbubble/ default.htm
  30. FDA Advisory Committee Briefing Document for the Cardiovascular and Renal Drugs Advisory Committee, 2008. Safety Considerations in the Development of Ultrasound Contrast Agents, 2008. Available at: www.fda.gov/ohrms/dockets/ac/08/briefing/2008-4369b1-01.pdf